PV8060
Pressure switch with IO-Link
PV-600-SEG14-UFRVG/US
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PV8060-00 — 21.02.2020 —
1 Sealing
Product characteristics
Output signal
switching signal; IO-Link; (configurable)
Measuring range 0...600 bar 0...8700 psi 0...60 MPa
Process connection
threaded connection G 1/4 external thread (DIN EN ISO 1179-2); internal thread M5
Application
Measuring element
metallic thin film cell
Application
for industrial applications
Media
liquids and gases
Medium temperature [°C]
-40...90
Pressure rating 1500 bar 21755 psi 150 MPa
Note on pressure rating static
Min. bursting pressure 2500 bar 36255 psi 250 MPa
Vacuum resistance [mbar]
-1000
Type of pressure
relative pressure
Electrical data
Operating voltage [V]
18...30 DC
Current consumption [mA]
< 15
Min. insulation resistance [MΩ]
100; (500 V DC)
Protection class
III
Reverse polarity protection
yes
Power-on delay time [s]
< 0.3
Inputs / outputs
Number of inputs and outputs
Number of digital outputs: 2
PV8060
Pressure switch with IO-Link
PV-600-SEG14-UFRVG/US
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PV8060-00 — 21.02.2020 —
Outputs
Total number of outputs
2
Output signal
switching signal; IO-Link; (configurable)
Electrical design
PNP/NPN
Number of digital outputs
2
Output function
normally open / normally closed; (parameterisable)
Max. voltage drop switching
output DC
[V]
2
Permanent current rating of
switching output DC
[mA]
100
Switching frequency DC [Hz]
< 130
Short-circuit protection
yes
Type of short-circuit
protection
pulsed
Overload protection
yes
Measuring/setting range
Measuring range 0...600 bar 0...8700 psi 0...60 MPa
Set point SP 6...600 bar 87...8702 psi 0.6...60 MPa
Reset point rP 3.1...597.1 bar 44...8660 psi 0.31...59.71 MPa
In steps of 0.1 bar 1 psi 0.01 MPa
Factory setting
SP1 = 150 bar rP1 = 138 bar ou1 = Hno;
SP2 = 450 bar rP2 = 438 bar ou2 = Hno;
dS1/dS2 = 0 ms dr1/dr2 = 0 ms
coF = 0 % P-n = PnP dAP= 60 ms
Temperature monitoring
Measuring range -40...90 °C -40...194 °F
Set point SP -38...90 °C -36.4...194 °F
Reset point rP -40...88 °C -40...190.4 °F
In steps of 0.1 °C 0.1 °F
Accuracy / deviations
Switch point accuracy
[% of the span]
< ± 0,5 (nach DIN EN 61298-2)
Repeatability [% of the span]
< ± 0,05; (with temperature fluctuations < 10 K)
Characteristics deviation
[% of the span]
< ± 0,5 (nach DIN EN 61298-2); (incl. drift when overtightened,
zero point and span error, non-linearity, hysteresis)
Linearity deviation
[% of the span]
< ± 0,1 (BFSL) / < ± 0,2 (LS)
Hysteresis deviation
[% of the span]
< ± 0,2
Long-term stability
[% of the span]
< ± 0,1; (per 6 months)
Temperature coefficient zero
point
[% of the span / 10 K]
< 0,1 (-25...90 °C) / < 0,2 (-40...-25 °C)
Temperature coefficient span
[% of the span / 10 K]
< 0,1 (-25...90 °C) / < 0,2 (-40...-25 °C)
PV8060
Pressure switch with IO-Link
PV-600-SEG14-UFRVG/US
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PV8060-00 — 21.02.2020 —
Temperature monitoring
Accuracy [K]
± 2 K + (0.1 x (ambient temperature - medium temperature))
Notes on the accuracy /
deviation
temperature range -10 to 80 °C
Response times
Response time [ms]
< 3
Software / programming
Parameter setting options
hysteresis / window; normally open / normally closed;
switching logic; switch-on/switch-off delay; Damping
Interfaces
Communication interface
IO-Link
Transmission type
COM2 (38,4 kBaud)
IO-Link revision
1.1
SDCI standard
IEC 61131-9
IO-Link device ID
1216 d / 00 04 c0 h
Profiles
Identification and Diagnosis (0x4000), Measurement Data Channel (0x800A)
SIO mode
yes
Required master port type
A
Process data analogue
5
Process data binary
2
Min. process cycle time [ms]
4.5
IO-Link resolution pressure [bar]
0.2
IO-Link resolution pressure [MPa]
0.02
IO-Link resolution
temperature
[K]
0.2
IO-Link process data
(cyclical)
function bit length
pressure 16
temperature 16
device status 4
binary switching information 2
IO-Link functions (acyclical)
application specific tag; internal temperature; operating hours counter;
switching cycles counter; Pressure peak counter; Temperature peak counter
Operating conditions
Ambient temperature [°C]
-40...90
Storage temperature [°C]
-40...100
Protection
IP 67; IP 69K
Tests / approvals
EMC DIN EN 61326-1
Shock resistance DIN EN 60068-2-27 500 g (1 ms)
Vibration resistance DIN EN 60068-2-6 20 g (10...2000 Hz)
UL approval
UL Approval no. J038
File number UL E174189
Pressure Equipment Directive
Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
PV8060
Pressure switch with IO-Link
PV-600-SEG14-UFRVG/US
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PV8060-00 — 21.02.2020 —
Mechanical data
Weight [g]
0.001
Materials
1.4542 (17-4 PH / 630); stainless steel (1.4404 / 316L); PEI
Materials (wetted parts)
stainless steel (1.4305 / 303); 1.4542 (17-4 PH / 630)
Min. pressure cycles
60 million; (at 1.2 times nominal pressure)
Tightening torque [Nm]
25...35; (recommended tightening torque; depends on lubrication, seal and pressure rating)
Process connection
threaded connection G 1/4 external thread (DIN EN ISO 1179-2); internal thread M5
Process connection sealing
FKM (to DIN 3869)
Restrictor element integrated
yes
Remarks
Remarks
BFSL = Best Fit Straight Line
LS = limit value setting
Pack quantity
1 pcs.
Electrical connection
Connector: 1 x M12
PV8060
Pressure switch with IO-Link
PV-600-SEG14-UFRVG/US
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PV8060-00 — 21.02.2020 —
Connection
OUT1 switching output pressure
IO-Link
OUT2 switching output pressure / temperature
colours to DIN EN 60947-5-2
Core colours :
BK = black
BN = brown
BU = blue
WH = white